  • for the system Altensteinstr. 23 a: all users must gain permission to work in the Biolabs
  • all users must get an introduction to the device by the responsible person
  • mark the time slot at which you want to do a measurement in the calendar, to avoid conflict with other user’s plan
  • DO NOT use solvents or other evaporating hazardous substances or gases inside the biosafety cabinet as they are no fume hoods and the HEPA-filtered air is circulating back in the room
  • DO NOT work with bacteria or viruses in the cell culture benches → for handling bacteria contact Elisa Quaas
  • Biosafety cabinets must be cleaned once a month with water and soap and wiped afterwards with desinfectant acccotding to the hygiene guide
  • Biosafety cabinets need to be tested by an authorized service once a year

...