- Gain permission to work in the Biolabs by follwoing the How-to-Start procedure.
- Get an introduction to the device by the responsible person (see above).
- DO NOT use the cell culture centrifuges for longer runs (>10min) → in case you need to centrifuge an extended time (>10min), make sure no one needs them for cell culture during this time. Otherwise use a centrifuge in a chemistry lab
- DO NOT use the cell culture centrifuges for other than biological samples, NO solvents etc. → For non-biological samples use a centrifuge in a chemistry lab
- Close centrifuge and tube lids during centrifugation
- Balance samples for centrifugation before using a balance and place them so that the rotor is in balance
- Wait after start until centrifuge reaches full speed → if the centrifuge shakes or is noisy stop immediately
- Open centrifuge after use and let lid open, so that no condensed water stay inside and cause corrosion
- Clean centrifuge after every use
- DO NOT use solvents or other evaporating hazardous substances → use instead a centrifuge in a chemistry lab
- Be cautious with rotors e.g. during cleaning, exchanging and store them safe
- Centrifuges must be cleaned once a month with water and soap and wiped afterwards with 70% ethanol
- Centrifuges need to be tested by an authorized service once a year
...