  • Gain permission to work in the Biolabs by follwoing the How-to-Start procedure.
  • Get an introduction to the device by the responsible person (see above).
  • DO NOT use the cell culture centrifuges for longer runs (>10min) → in case you need to centrifuge an extended time (>10min), make sure no one needs them for cell culture during this time. Otherwise use a centrifuge in a chemistry lab
  • DO NOT use the cell culture centrifuges for other than biological samples, NO solvents etc. → For non-biological samples use a centrifuge in a chemistry lab
  • Close centrifuge and tube lids during centrifugation
  • Balance samples for centrifugation before using a balance and place them so that the rotor is in balance
  • Wait after start until centrifuge reaches full speed → if the centrifuge shakes or is noisy stop immediately
  • Open centrifuge after use and let lid open, so that no condensed water stay inside and cause corrosion
  • Clean centrifuge after every use
  • DO NOT use solvents or other evaporating hazardous substances → use instead a centrifuge in a chemistry lab
  • Be cautious with rotors e.g. during cleaning, exchanging and store them safe
  • Centrifuges must be cleaned once a month with water and soap and wiped afterwards with 70% ethanol
  • Centrifuges need to be tested by an authorized service once a year

...

Safety Operation Instruction:

 see BAs Devices