Location: Altensteinstr. 23a, Biolabs 114, 114.5, 114.6, 114.7, 115, 115.5, 115.6, 115.7 (S2 Biolabs) & 027.1 (S2 OM)
Responsible Persons:
Elisa Quaas, phone: 63652, mail: eqaas@zedat.fu-berlin.de
Johanna Scholz, phone: 71087, mail: johanna.scholz1@fu-berlin.de
Katharina Achazi, phone: 59145, katharina.achazi@fu-berlin.de
Stefanie Wedepohl, phone 58721, stefanie.wedepohl@fu-berlin.de
Requirements:
- Gain permission to work in the Biolabs by follwoing the How-to-Start procedure.
- Get an introduction to the device by the responsible person (see above).
- DO NOT use the cell culture centrifuges for longer runs (>10min) → in case you need to centrifuge an extended time (>10min), make sure no one needs them for cell culture during this time. Otherwise use a centrifuge in a chemistry lab
- DO NOT use the cell culture centrifuges for other than biological samples, NO solvents etc. → For non-biological samples use a centrifuge in a chemistry lab
- Close centrifuge and tube lids during centrifugation
- Balance samples for centrifugation before using a balance and place them so that the rotor is in balance
- Wait after start until centrifuge reaches full speed → if the centrifuge shakes or is noisy stop immediately
- Open centrifuge after use and let lid open, so that no condensed water stay inside and cause corrosion
- Clean centrifuge after every use
- DO NOT use solvents or other evaporating hazardous substances → use instead a centrifuge in a chemistry lab
- Be cautious with rotors e.g. during cleaning, exchanging and store them safe
- Centrifuges must be cleaned once a month with water and soap and wiped afterwards with 70% ethanol
- Centrifuges need to be tested by an authorized service once a year
Application/Technical Details:
Centrifugation of samples for 1.5ml to 50ml Tubes and well plates
Instruction/Manual
Serial No.:
42095369 (Heraeus Megafuge 8R, Takustr. 6, K004)
41696601 (Heraeus Megafuge 8R, Arnimallee 22, E002)
721114110508 (Heraeus Megafuge 8, Arnimallee 22, E004)
FUB Inventory No.:
72000057270 (Heraeus Megafuge 8R, Takustr. 6, K004)
3136385 Charite (Heraeus Megafuge 8R, Arnimallee 22, E002)
72000032270 (Heraeus Megafuge 8, Arnimallee 22, E004)