Location:
Altensteinstr. 23a, Biolabs 114.XX, 115.XX, 027.1
Responsible Person:
Application/Technical Details
Requirements:
- All users must gain permission to work in the Biolabs and access to the “Zellbio-users” network drive by attending the Biosafety instructions (--> Katharina Achazi)
- All users must get an introduction to the device from the resposnible person (OpenIRIS) before use based on the operation manuals (see below) documented in the device log book
- Only use the tubes that are compatible with the rotors we have and follow the tube protocol for filling and operating
- ech run needs to be documented in the device log book
- In case of spillage, open rotor in a sterile bench and clean centrifuge
- DO NOT use organic solvents or other evaporating hazardous substances → use a centrifuge in a chemistry lab instead
- close centrifuge and tube lids during centrifugation
- precisely weigh samples and place them so that the rotor is in balance
- wait after start until centrifuge reaches full speed → if the centrifuge shakes or is noisy stop immediately
- open centrifuge after use and leave lid open, so that no condensed water stay inside and cause corrosion
- clean centrifuge after every use
- be cautious with rotors e.g. during cleaning, exchanging and store them safe
- centrifuges must be cleaned once a month with water and soap and wiped afterwards with 70% ethanol
- centrifuges need to be tested by an authorized service once a year
Instruction/Manual
Serial No.:
FUB Inventory No.:
Safety Operation Instruction (German):
{include} konnte nicht gerendert werden Die eingeschlossene Seite konnte nicht gefunden werden.