Versionen im Vergleich

Schlüssel

  • Diese Zeile wurde hinzugefügt.
  • Diese Zeile wurde entfernt.
  • Formatierung wurde geändert.

For Training please contact: Christian.Halbig@fu-Berlin.de

Notes: Cantilever for basic measurements are provided by NanoFAB.

Setup History

  • 2016: Purchased with cAFM and Kalvin Probe module.
  • 2022: Relocated to SupraFAB

Official Documents and Operation

Quick Operation Guide

  • Download Document (PDF) (Word)
    • Device Handling

    • Contact Mode

    • AC (Intermittent) Mode

    • Conductive AFM
    • Liquid AFM
  • Download Summary of Steps (PDF) (Word)

Tips for Cantilever

  • https://www.nanoandmore.com/
    • AC Mode (AIR):
      •  BudgetSensors Tap300: -G (uncoated), Al-G (aluminium coated), GD-G or GB-G (gold coated)
        • they are quite good standard tips (tip radius <10nm). Choose gold coated for liquid-AFM.
        • 10 pcs / ~200 €
      • Nanosensors PPP-NCx (Point Probe Plus - Non-Contact), NCH (higher frequency), NCL (lower frequency), NCR (aluminium coated), NCAuD (gold coated)
        • Slightly better in performance, but also more expensive.
        • 10 pcs / ~330 €
    • AC Mode (LIQUID):
      • BudgetSensors Multi75GD-G or Multi75GB-G (both gold coated)
        • Recommended for basic liquid-AFM using AC mode.
        • 10 pcs / ~260 €
    • Contact Mode:
      • AIR: BudgetSensors Contact-G (uncoated) or ContAl-G (aluminium coated) 10 pcs / ~200 €
      • LIQUID: BudgetSensors ContGD-G (gold coated) 10 pcs / ~260 €
      • Conductive-AFM:
        • PPP-ContPt (Nanosensors) orContPt (NanoWorld) 10 pcs / ~420-460 €
        • ElektriCont-G (BudgetSensors) 10 pcs / ~240 €

Downloads

SoftwareDescriptionLink
GwyddionFreeware to process recorded SPM data. Available for Windows, Linux and Mac. http://gwyddion.net/
WSxMSoftware to process recorded AFM data. Free for non-commercial usage. Windows only. http://www.wsxm.eu/
JPK SPM ProcessingOfficial JPK Software (v. 6.0.65) for processing your data. Windows only<ask cleanroom management>

DescriptionLink
Artifacts in AFMA nice summary of artifacts in AFM.

View file
nameAFM_Bild-Artefakte.pdf
height250

Artifacts in AFM

F. Gołek, P. Mazur, Z. Ryszka, S. Zuber: "Artifacts in AFM"

https://doi.org/10.1016/j.apsusc.2014.01.149
Application Note of NT-MDT:

T. Füzik, P. Ulbrich, T. Ruml: "Imaging of biological smaples in liquid environment"

Institute of Chemical Technology Prague, Department of Biochemistry and Microbiology

View file
name080_Imaging_of_biological_samples_in_liquid__A4_en.pdf
height250

Calibration of AFM cantilever

The following link are useful in order to calibrate the spring constant of a cantilever. The

first method is the one I am more partial. The second was considered the gold standard

but the above method has gained respect.

http://www.physics.uwo.ca/~hutter/calibration/afmcal.html

http://www.asylumresearch.com/Applications/CantileverCal/AddedMass.pdf

...