  • All users must gain permission to work in the Biolabs and access to the “Zellbio-users” network drive by attending the Biosafety instructions (--> Katharina Achazi)
  • All users must get an introduction to the device from the resposnible person (OpenIRIS) before use based on the operation manula manuals (see below) documented in the device log book
  • Only use the tubes that are compatible with the rotors we have and follow the tube protocol for filling and operating
  • ech run needs to be documented in the device log book
  • In case of spillage, open rotor in a sterile becnh bench and clean centrifuge
  • DO NOT use the cell culture centrifuges for longer runs (>10min) → in case you need to centrifuge an extended time (>10min), make sure no one needs them for cell culture during this time. Otherwise use a centrifuge in a chemistry labDO NOT use the cell culture centrifuges for other than biological samples, NO solvents etc. → For non-biological samples organic solvents or other evaporating hazardous substances → use a centrifuge in a chemistry lab instead
  • close centrifuge and tube lids during centrifugation
  • balance samples for centrifugation before using a balance precisely weigh samples and place them so that the rotor is in balance
  • wait after start until centrifuge reaches full speed → if the centrifuge shakes or is noisy stop immediately
  • open centrifuge after use and let leave lid open, so that no condensed water stay inside and cause corrosion
  • clean centrifuge after every use
  • DO NOT use solvents or other evaporating hazardous substances → use instead a centrifuge in a chemistry lab
  • be cautious with rotors e.g. during cleaning, exchanging and store them safe
  • centrifuges must be cleaned once a month with water and soap and wiped afterwards with 70% ethanol
  • centrifuges need to be tested by an authorized service once a year

...