Optima MAX-XPImage Added

Location:

Altensteinstr. 23a, Biolabs 114.7


Responsible Person:

Application/Technical Details

Requirements:

persons & booking: https://fub.openiris.io/timeline/73597637-99d1-41ac-b346-839f70103473

Detailed information and protocols how to use OpenIris you find here.


Application/technical details:

Centrifugation of materials at very high speed.


Requirements:

  • Gain All users must gain permission to work in the Biolabs and by follwoing the How-to-Start procedure.
  • Get access to the “ZellbioBCP-users” storage network drive by attending the Biosafety instructions (--> Katharina Achazi)suprafab-analytics to save measuremnt data.
  • Book the device using OpenIRIS (protocls see here).
  • Get All users must get an introduction to the device from by the resposnible responsible person (OpenIRISsee above) before use based on the operation manuals (see below) documented in the device log book
  • Only use the tubes that are compatible with the rotors we have and follow the tube protocol for filling and operating
  • ech Ech run needs to be documented in the device log book
  • In case of spillage, open rotor in a sterile bench and clean centrifuge
  • DO NOT use organic solvents or other evaporating hazardous substances → use a centrifuge in a chemistry lab instead
  • close Close centrifuge and tube lids during centrifugation
  • precisely Precisely weigh samples and place them so that the rotor is in balance
  • wait Wait after start until centrifuge reaches full speed → if the centrifuge shakes or is noisy stop immediately
  • open Open centrifuge after use and leave lid open, so that no condensed water stay inside and cause corrosion
  • clean Clean centrifuge after every use
  • be Be cautious with rotors e.g. during cleaning, exchanging and store them safe
  • centrifuges Centrifuges must be cleaned once a month with water and soap and wiped afterwards with 70% ethanol
  • centrifuges Centrifuges need to be tested by an authorized service once a year

...