Optima MAX-XP

Location: Altensteinstr. 23a, Biolabs 114.7

Responsible Persons & Booking: https://fub.openiris.io/timeline/73597637-99d1-41ac-b346-839f70103473

Detailed information and protocols how to use OpenIris you find here.

Application/technical details:

Centrifugation of materials at very high speed.

Requirements:

  • Gain permission to work in the Biolabs by follwoing the How-to-Start procedure.
  • Get access to the BCP-storage network drive suprafab-analytics to save measuremnt data.
  • Book the device using OpenIRIS (protocls see here).
  • Get an introduction to the device by the responsible person (see above) before use based on the operation manuals (see below) documented in the device log book
  • Only use the tubes that are compatible with the rotors we have and follow the tube protocol for filling and operating
  • Ech run needs to be documented in the device log book
  • In case of spillage, open rotor in a sterile bench and clean centrifuge
  • DO NOT use organic solvents or other evaporating hazardous substances → use a centrifuge in a chemistry lab instead
  • Close centrifuge and tube lids during centrifugation
  • Precisely weigh samples and place them so that the rotor is in balance
  • Wait after start until centrifuge reaches full speed → if the centrifuge shakes or is noisy stop immediately
  • Open centrifuge after use and leave lid open, so that no condensed water stay inside and cause corrosion
  • Clean centrifuge after every use
  • Be cautious with rotors e.g. during cleaning, exchanging and store them safe
  • Centrifuges must be cleaned once a month with water and soap and wiped afterwards with 70% ethanol
  • Centrifuges need to be tested by an authorized service once a year

Instruction/Manual

View file
name393552AM.pdf
height250
View file
name393812AK.pdf
height250

View file
namerotors and tubes.pdf
height250
View file
nameMLA-50.pdf
height250
View file
nameTLA-110.pdf
height250

Serial No.: CTZ21G011

FUB Inventory No.: 7200009337

Safety Operation Instruction (German):

Seite einschließen
BA Ultrazentrifugen