Location: Altensteinstr. 23a, Biolabs 114.7
Responsible Persons & Booking: https://fub.openiris.io/timeline/73597637-99d1-41ac-b346-839f70103473
Detailed information and protocols how to use OpenIris you find here.
Application/technical details:
Centrifugation of materials at very high speed.
Requirements:
- Gain permission to work in the Biolabs by follwoing the How-to-Start procedure.
- Get access to the BCP-storage network drive suprafab-analytics to save measuremnt data.
- Book the device using OpenIRIS (protocls see here).
- Get an introduction to the device by the responsible person (see above) before use based on the operation manuals (see below) documented in the device log book
- Only use the tubes that are compatible with the rotors we have and follow the tube protocol for filling and operating
- Ech run needs to be documented in the device log book
- In case of spillage, open rotor in a sterile bench and clean centrifuge
- DO NOT use organic solvents or other evaporating hazardous substances → use a centrifuge in a chemistry lab instead
- Close centrifuge and tube lids during centrifugation
- Precisely weigh samples and place them so that the rotor is in balance
- Wait after start until centrifuge reaches full speed → if the centrifuge shakes or is noisy stop immediately
- Open centrifuge after use and leave lid open, so that no condensed water stay inside and cause corrosion
- Clean centrifuge after every use
- Be cautious with rotors e.g. during cleaning, exchanging and store them safe
- Centrifuges must be cleaned once a month with water and soap and wiped afterwards with 70% ethanol
- Centrifuges need to be tested by an authorized service once a year
Instruction/Manual
Serial No.: CTZ21G011
FUB Inventory No.: 7200009337
Safety Operation Instruction:
see Devices BAs